The effect of ADMIRE-based low-tube voltage CT scanning on radiation dose and image quality of the upper abdomen |
CHEN Chunmiao, LIN Guihan, WU Xulu, LU Chenying, YANG Hongyuan, CHEN Xiao, JI Jiansong, HU Xianghua |
Key Laboratory of Imaging Diagnosis and Minimally Invasive Intervention Research, Department of Radiology, Lishui Central Hospital, Lishui 323000, China |
|
Cite this article: |
CHEN Chunmiao,LIN Guihan,WU Xulu, et al. The effect of ADMIRE-based low-tube voltage CT scanning on radiation dose and image quality of the upper abdomen[J]. JOURNAL OF WEZHOU MEDICAL UNIVERSITY, 2020, 50(1): 56-59,64.
|
|
Abstract Objective: To explore the effect of low-tube voltage CT scanning on radiation dose and image quality of the upper abdomen based on advanced model iterative reconstruction algorithm (ADMIRE). Methods: Totally 200 patients who had undergone upper abdomen scanning in Lishui Central Hospital from November 2016 to September 2017 were enrolled and randomly assigned to the control group (n=50) and the experimental group (n=150), who were all treated with ADMIRE with the tube voltage of 120 kV in the control group. Meanwhile, the experimental group was divided as low-tube voltage A, B, and C according to the body mass index (BMI) with the tube voltage of 80 kV, 90 kV and 100 kV respectively. Subjective scores of images, background noise standard deviation (SD), liver signal-to-noise ratio (SNR), liver contrast-to-noise ratio (CNR) and effective radiation dose (ED) were compared between each group. Results: There were no significant differences between the control group and all the low-tube voltage groups (P>0.05) in age, gender or CT values of erector spinal and the image quality scores, with all images meeting the clinical diagnostic requirements, but there were significant differences (P<0.05) in BMI and liver CT. The BMI of the low-tube voltage groups A and B was significantly lower than that of the control group, while that of the low-tube voltage group C was significantly higher than the control group. Liver CT value of the low-tube voltage group A was significantly higher than the control group. The background noise SD of the images of the low-tube voltage groups A, B, and C were significantly lower than those of the control group, while the liver SNR and liver CNR were significantly higher than those of the control group. Meanwhile, the ED of the low tube voltages A, B and C was significantly lower than that of the control group (P<0.05). Conclusion: When applying the third-generation dual-source CT to the upper abdominal scanning, ADMIRE-based low-tube voltage scanning mode can significantly reduce the radiation dose and ensure the image quality.
|
Received: 16 April 2019
|
|
|
|
|
[1] 石清磊, 赵红梅, 张玲, 等. 自适应统计迭代重建算法对腹部CT扫描中图像质量和辐射剂量影响的模体研究[J]. 中华放射学杂志, 2013, 47(4): 326-329.
[2] PIECHOWIAK E I, PETER J F, KLEB B, et al. Intravenous iodinated contrast agents amplify DNA radiation damage at CT[J]. Radiology, 2015, 275(3): 692-697.
[3] 中华医学会放射学分会质量管理与安全管理学组. CT辐射剂量诊断参考水平专家共识[J]. 中华放射学杂志, 2017, 51(11): 817-822.
[4] 陈永华, 傅钢泽, 陈聪, 等. 低管电压CTA在诊断急性下肢创伤合并血管损伤中的应用价值[J]. 温州医科大学学报, 2018, 48(9): 667-671.
[5] 陈永华, 傅钢泽, 潘克华, 等. 低管电压联合迭代重建技术对自体上肢动静脉内瘘CTA成像质量及辐射剂量的影响[J]. 温州医科大学学报, 2019, 49(4): 289-293, 297.
[6] SINGH S, KALRA M K, GILMAN M D, et al. Adaptive statistical iterative reconstruction technique for radiation dose reduction in chest CT: a pilot study[J]. Radiology, 2011, 259(2): 565-573.
[7] 刘晓冬, 刘爱连, 刘静红, 等. 基于多模型的迭代重建技术对低剂量CT结肠成像图像质量的影响[J]. 中华放射医学与防护杂志, 2018, 38(4): 311-316.
[8] 王苑丁, 姚新宇, 苏壮志, 等. 应用高级迭低重建改善低剂量冠状动脉CT血管成像图像质量的研究[J]. 医学影像学杂志, 2017, 27(5): 819-824.
[9] 唐坤, 曹国全, 李瑞, 等. 低管电压腹部CT扫描对图像质量及辐射剂量影响的体模实验[J]. 中国医学影像技术, 2012, 28(4): 800-804.
[10] 袁肖娜, 高知玲, 马文东, 等. 对比分析容积CT剂量指数与体型特异性的剂量评估在估算腹部CT扫描辐射剂量中的差异[J]. 中华放射医学与防护杂志, 2016, 36(1): 74-77.
[11] 李锋坦, 李东, 张云亭. 管电压对CT值测量、辐射剂量及图像质量影响的模型研究[J]. 中华放射学杂志, 2013, 47(5): 458-461.
[12] GORDIC S, DESBIOLLES L, STOLZMANN P, et al. Advanced modelled iterative reconstruction for abdominal CT: qualitative and quantitative evaluation[J]. Clin Radiol, 2014, 69(12): 497-504.
[13] 许伟雄, 黄旭璇, 林顺发, 等. 腹部动脉能谱CT混合能量与单能量图像的比较研究[J]. 中国CT和MRI杂志, 2014, 12(8): 91-94.
[14] 潘元威, 周志刚, 高剑波, 等. 能谱CT智能匹配技术在不同体质量指数患者腹部CT扫描中的应用[J]. 临床放射学杂志, 2017, 36(7): 1034-1037.
[15] 霍丙胜, 姬国敏, 杨海庆. 肝脏能谱CT与常规CT辐射剂量和图像质量对比研究[J]. 河北医药, 2015, 37(18): 2779-2781. |
|
|
|