The third-generation dual-source CT+flash mode scan combined with chest circumference index to reduce the radiation dose and iodine intake of coronary CTA |
WANG Hailin, CHEN Minjiang, WENG Qiaoyou, LU Chenying, CHENG Shufang, YE Miaoqing, LIN Guihan, JI Jiansong |
Department of Radiology, Lishui Central Hospital, Lishui 323000, China |
|
Cite this article: |
WANG Hailin,CHEN Minjiang,WENG Qiaoyou, et al. The third-generation dual-source CT+flash mode scan combined with chest circumference index to reduce the radiation dose and iodine intake of coronary CTA[J]. JOURNAL OF WEZHOU MEDICAL UNIVERSITY, 2020, 50(1): 36-40.
|
|
Abstract Objective: To explore the clinical application value of third-generation dual-source CT+flash mode scan combined with chest circumference index to reduce the radiation dose and iodine intake of coronary artery CT angiography (CTA). Methods: A total of 160 patients with clinically suspected coronary heart disease requiring coronary CTA examination were continuously selected and examined by trube-flash large pitch scan combined with iterative reconstruction technology. Tube current automatic adjustment technology care Dose 4D was operated. The control group included 60 cases, with tube voltage=120 kV, and the remaining 100 cases were in the low-dose group, which were sub-divided into groups A, B, and C according to the parameters of the chest circumference. Among them, the chest circumference of group A was≤85 cm, the tube voltage was 70 kV; group B was 85-90 cm, tube voltage=80 kV; group C was≥95 cm, tube voltage=90 kV. One-way analysis of variance was used to compare gender, age, BMI, chest circumference, tube voltage, tube current, right coronary CT value, SD value, SNR, CTD, DLP, radiation dose in four groups of patients; the comparison of subjective image quality score was performed by a nonparametric rank sum test. Results: The coronary image quality scores of the conventional group and the low-dose group (group A, group B, and group C) were 3.26±0.23, 3.18±0.19, 3.23±0.31, and 3.21±0.25, respectively. There was no significant difference between the groups (P>0.05), and the image quality met the requirements of clinical diagnosis. There were no significant differences in age, gender and heart rate between the 4 groups (P>0.05). The CT values of the four groups of right coronary artery were (425.94±69.59)Hu, (505.23±41.41) Hu, (463.83±59.68) Hu and (444.70±70.36) Hu, respectively, and there was statistical difference among the low-dose groups and the conventional group (P<0.01). The radiation doses were (1.03±0.15) mSv, (0.36±0.09) mSv, (0.57±0.05) mSv and (0.80±0.10) mSv (P>0.05); the iodine intakes were 18.50 g, 9.45 g, 12.80 g, and 14.40 g, respectively. The radiation dose and iodine intake of the low-dose groups were lower than the conventional group (P<0.01). Conclusion: The third-generation dual-source CT+flash mode scanning combined with chest circumference index can be applied to adjust the tube voltage and meanwhile the contrast agent with low iodine concentration can be taken to adjust the rate and total amount of contrast agent, thus effectively reducing the radiation dose and iodine intake while the image quality is ensured.
|
Received: 30 January 2019
|
|
|
|
|
[1] FRUSH D P, APPLEGATE K. Computed tomography and radiaition: under standing the issues[J]. J Am Coll Radiol, 2004, 1(2): 113-119.
[2] 张丽, 于红, 刘士远, 等. 迭代重建技术对低剂量肺部平扫CT图像质量的影响[J]. 中华放射学杂志, 2013, 47(4): 316-320.
[3] 陈淮, 曾庆思, 郑劲平, 等. 低剂量螺旋CT定量成像技术对吸烟患者肺功能的研究[J]. 中华放射学杂志, 2012, 46(5): 405-409.
[4] ZHAO Y, TAO Z, XU Z, et al. Toxic effects of a high dose of non-ionic iodinated contrast media on renal glomerular and aortic endothe lial cells in aged rats in vivo[J]. Toxicol Lett, 2011, 202(3): 253-260
[5] 高建华, 王贵生, 郑静晨, 等. 体重指数在64层螺旋CT心脏扫描X线剂量管理中的应用研究[J]. 中华放射学杂志, 2008, 42(8): 877-882.
[6] SZUCS-FARKAS Z, SCHALLER C, BENSLER S, et al. Detection of pulmonary emboli with CT angiography at reduced radiation exposure and contrast material volume: comparison of 80 kVp and 120 kVp protocols in a matched cohort[J]. Invest Radiol, 2009, 44(12): 793-799.
[7] LELL M, HINKMANN F, ANDERS K, et al. High-pitch electrocardiogram-triggered computed tomography of the chest: initial results[J]. Invest Radiol, 2009, 44(11): 728-733.
[8] 王怡宁, 曹剑, 孔令燕, 等. 高灵敏度探测器结合迭代重建在低管电压冠状动脉CT血管成像的价值[J]. 中华放射学杂志, 2014, 48(2): 109-113.
[9] 王海林, 纪建松, 卢陈英, 等. 胸围调节管电压技术冠状动脉CT血管成像的图像质量和辐射剂量[J]. 中华放射学杂志, 2016, 50(4): 284-288.
[10] ZHANG C, ZHANG Z, YAN Z, et a1. 320-row CT coronary angiography: effect of 100-kV tube voltages on image quality, contrast volume, and radiation dose[J]. Int J Cardiovasc Imaging, 201l, 27(7): 1059-1068. |
|
|
|