Establishment of modified microcolumn gel method for detection of ABO variant blood type |
ZHU Suiyong, JIN Fangsi, SHI Shunqiu, QIU Xiaole, ZHENG Tingting |
Department of Blood Transfusion, the Second Affiliated Hospital & Yuying Children’s Hospital of Wenzhou Medical University, Wenzhou 325027, China |
|
Cite this article: |
ZHU Suiyong,JIN Fangsi,SHI Shunqiu, et al. Establishment of modified microcolumn gel method for detection of ABO variant blood type[J]. JOURNAL OF WEZHOU MEDICAL UNIVERSITY, 2019, 49(2): 104-107.
|
|
Abstract Objective: To establish a modified microcolumn gel method for the detection of ABO variant blood type. Methods: From November 2016 to May 2018, 12 patients with ABO blood group agglutination intensity change or inconsistent positive and negative stereotypes were selected from the Second Affiliated Hospital &
Yuying Children’s Hospital of Wenzhou Medical University, Wenzhou People’s Hospital and Rui’an People’s Hospital. The ABO blood type of the specimens was confirmed by blood type serogy test and blood type gene test combined with clinical manifestations. The known blood group specimens were detected by microcolumn gel method. A modified microcolumn gel method was established by using 1% O, A3 subtype and A antigen weakened erythrocytes to adjust the way of centrifugation of microcolumn gel method, which was used to detect all the samples. Results: Of the 12 cases, there was 1 case of A3 subtype, 3 cases of B(A) type, 3 cases of Mumbai type and 5 cases of ABO blood group antigen with weak expression. The sensitivity of microcolumn gel method was similar to that of saline tube test. Only 2 of 12 specimens were detected, with the detection rate being 2/12. The centrifugation method of the modified microcolumn gel method was 900 r/min centrifugation 2 min, 30 min incubated at 4 °C and 1 800 r/min centrifugation 3 min. Only 1 case of the 12 specimens could not be detected by the modified microcolumn gel method, the detection rate being 11/12. Conclusion: The modified microcolumn gel method established in this study improves the detection sensitivity by changing the centrifuge mode and increased incubation time and incubation temperature of the microcolumn gel blood type card. It is suitable for screening ABO variant blood groups.
|
|
|
|
|
|
[1] HOSOI E. Biological and clinical aspects of ABO blood group system[J]. J Med Invest, 2008, 55(3-4): 174-182.
[2] YAZER M H, TRIULZI D J. Immune hemolysis following ABO-mismatched stem cell or solid organ transplantation.[J]. Curr Opin Hematol, 2007, 14(6): 664-670.
[3] 方建华, 刘玉振, 温涛. 国产单克隆抗体试剂漏检Ax亚型1例[J]. 中国输血杂志, 2004, 17(2): 119-119.
[4] 迟晓云, 闫萍, 韩斌, 等. 微柱凝胶卡漏检B_w03亚型分析——附2例报告[J]. 中国输血杂志, 2016, 29(10): 1147-1150.
[5] 何卫社, 刘乐霞, 王德平, 等. 全自动血型仪漏检4例B(A)亚型原因分析[J]. 临床血液学杂志, 2016, 29(8): 642-644.
[6] 刘乐霞, 肖玮, 梁俊杰, 等. 国产单克隆血型试剂漏检B(A)血型分析[J]. 临床输血与检验, 2012, 14(4): 333-334.
[7] 李勇, 杨贵贞. 人类红细胞血型学实用理论与实验技术[M]. 北京: 中国科学技术出版社, 1999, 267.
[8] 刘秋菊, 朱碎永, 林甲进. 类孟买型血清学特点及后续输血策略的研究[J]. 中国卫生检验杂志, 2017, 27(9): 1276-1278.
[9] 林晨蓓, 王秀娣, 朱碎永, 等. A3亚型血型的血清学和分子生物学鉴定分析[J]. 中国卫生检验杂志, 2018, 28(7): 801-802, 829.
[10] TITLESTAD K, GEORGSEN J, ANDERSEN H, et al. Detection of irregular red cell antibodies: more than 3 years of experience with a gel technique and pooled screening cells [J]. Vox Sang, 1997, 73(4): 246-251.
[11] LANGSTON M M, PROCTER J L, CIPOLONE K M, et al. Evaluation of the gel system for ABO grouping and D typ-ing[J]. Transfusion, 1999, 39(3): 300-305.
[12] 于洋, 冯倩, 林子林, 等. 全自动微柱凝胶技术在ABO、
RhD血型鉴定中的应用研究[J]. 北京医学, 2008, 30(3): 165-167.
[13] 于海忠. 微柱凝胶法在血型鉴定和交叉配血中的应用[J]. 临床和实验医学杂志, 2010, 9(6): 453-454.
[14] 李毅, 佟力, 董伟群, 等. 微柱凝胶卡法在血型检测中的应用[J]. 昆明医学院学报, 2010, 31(1): 118-119, 126.
[15] 肖倩, 辛荣传, 周益强, 等. 微柱凝胶法检测ABO疑难血型的临床应用及方法学探讨[J]. 中国实验诊断学, 2009, 13(7): 932-934.
[16] 周小玉, 许进明, 华岚, 等. 微柱凝胶法鉴定ABO血型正反定型不一致原因分析[J]. 临床输血与检验, 2007, 9(4): 366-367.
[17] 邹纬, 丁秋兰, 王学锋. 血型卡在血型鉴定中的局限性[J]. 诊断学理论与实践, 2005, 4(5): 410-411.
[18] 郭伟鹏, 周吉霞, 吐尔洪·克维尔, 等. 微柱凝胶法鉴定32例ABO亚型结果分析[J]. 临床血液学杂志(输血与检验), 2015, 28(4): 723-724.
[19] 钟胜英. 手工微柱凝胶免疫检验法在ABO、RhD血型抗原鉴定中的应用效果[J]. 实用中西医结合临床, 2017, 17(5): 4-5.
[20] 林海芬, 谢作听, 王春香. 改良微柱凝胶法在鉴定ABO血型抗原明显减弱者中的应用[J]. 临床输血与检验, 2004, 6(4): 267-269. |
|
|
|